Description
The EL037 has 4 input channels when used with the EL005 logger (3 channels are available when used with the EL008), each channel can be be configured independently (using jumpers) to measure ±2.5 V, ±10 V or 4–20 mA. It can also route an external power supply (perhaps 12 V or 24 V, depending on the requirements of the input devices) to each of the input devices. Connections to the input channels are via screw terminals.
Range | Accuracy | Resolution |
±2.5 V | -16 bits + sign | ±0.2% |
±10 V | -16 bits + sign | ±0.4% |
4-20mA | -16 bits + sign | ±0.3% |